In this episode Shahriar repairs an HP 8562B 22GHz spectrum analyzer. This spectrum analyzer has two input bands. The lower band which operates from 1kHz – 2.9GHz shows does not operate correctly and does not display any signal. The internal construction as well as the operation of the unit is explained along with a close look at LO distribution, input attenuator, dual-mixer and various filters.
During the debugging procedure it is also discovered that the Second Mixer is also damaged in this unit. A replacement unit was purchased however a replacement unit for the main first converter is too expensive to replace. As an alternative, a Mini-Circuit mixer is used as a replacement. The LO signal is routed to the external mixer via a microwave switch. In this configuration the LO is automatically switched to the correct mixer for each frequency band of operation. The unit is then verified for correct operation.
The Signal Path Blog
Mdp.lt is not the owner of this text/video/image/photo content, the real source of content is Youtube.com and user declared in this page publication as Youtube.com user, if you have any question about video removal, what was shared by open community, please contact Youtube.com directly or report bad/not working video links directly to video owner on Youtube.com. Removed video from Youtube.com will also be removed from here.